Estimation of right ventricular systolic pressure by Doppler echo is from: a) Velocity of tricuspid regurgitation jet b) Velocity of pulmonary regurgitation jet c) Area of tricuspid regurgitation
Measures useful in preventing interaction between implanted pacemaker and electrocautery: a) Using bipolar cautery b) Placing the indifferent plate electrode in such a way that the current path
Function of magnetic reed switch in an ICD (implantable cardioverter-defibrillator): a) Changes pacing mode to asynchronous (e.g. VVI to VOO) b) Suspends all tachyarrhythmia therapies c) None of
Enhanced tendency for arrhythmias in pregnancy In general there is an enhancement of tendency for arrhythmias in pregnancy. Isolated ventricular ectopic beats are probably the commonest arrhythmia in
Right ventricular apical pacing and mitral regurgitation Right ventricular apical pacing may sometimes cause mitral regurgitation or worsen a pre-existing mitral regurgitation. Several mechanisms have been considered for
Electrosurgery and pacemakers Electrosurgical equipment use radiofrequency energy in the range of 0.1 to 5 MHz. This can cause electromagnetic interference (EMI) with pacemakers and other cardiac implantable
MRI conditional pacemakers Conventionally MRI (magnetic resonance imaging) is contraindicated in those with implanted pacemakers due to possibility of electromagnetic interference (EMI) with pacemaker function. MRI-conditional means that